Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Sub Yokogawa uR100 (1 year) Temperature CAL VER /5500 DATE: 15-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 47 NUMBER OF LINES: 178 CONFIGURATION: Fluke 5500A ============================================================================= # # Requires: # Channel # (1-3) in M[1] # Temperature range # (10-16) in M[3] # # Compatibility: # 5500/CAL 4.23 or MET/CAL 4.23 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P J F W 1.002 ASK+ K X 1.003 MATH MEM = M[1] 1.004 HEAD CHANNEL [MEM] THERMOCOUPLE 1.005 RSLT =CHANNEL [MEM] THERMOCOUPLE 1.006 DISP Connect the 5500A TC terminals to Ch[MEM] as follows: 1.006 DISP [32] 5500A TC + to A TOP INPUT TERMINAL 1.006 DISP [32] 5500A TC - to B BOTTOM INPUT TERMINAL 1.007 DISP Press the keyboard SET until RNG code is displayed. 1.008 DISP 1.008 DISP Set the CHANNEL NUMBER to [MEM]. 1.008 DISP SET CODE to A (shift 1) 1.008 DISP RANGE CODE to [MEM1] 1.008 DISP 1.008 DISP Press the key to store the data. 1.009 JMPL TYPE_E_AND_T ZCMPI(MEM2, "E") 1.010 JMPL TYPE_E_AND_T ZCMPI(MEM2, "T") # Type R, S, B, K, J, and N 1.011 DISP Set the SCALE to 0degC - 1000degC as follows: 1.011 DISP 1.011 DISP Press <0> key 5 times. 1.011 DISP Press to store scale minimum to 0degC 1.011 DISP 1.011 DISP Press <1> <0> <0> <0> <0> 1.011 DISP Press to store scale maximum to 1000degC 1.012 JMPL THERMOCOUPLE_TYPE_R ZCMPI(MEM2, "R") 1.013 JMPL THERMOCOUPLE_TYPE_S ZCMPI(MEM2, "S") 1.014 JMPL THERMOCOUPLE_TYPE_B ZCMPI(MEM2, "B") 1.015 JMPL THERMOCOUPLE_TYPE_K ZCMPI(MEM2, "K") 1.016 JMPL THERMOCOUPLE_TYPE_J ZCMPI(MEM2, "J") 1.017 JMPL THERMOCOUPLE_TYPE_N ZCMPI(MEM2, "N") 1.018 LABEL TYPE_E_AND_T 1.019 DISP Set the SCALE to -100degC - 400degC as follows: 1.019 DISP 1.019 DISP Press <-> <1> <0> <0> <0> 1.019 DISP Press to store scale minimum to -100degC 1.019 DISP 1.019 DISP Press <0> <4> <0> <0> <0> 1.019 DISP Press to store scale maximum to 400degC 1.020 JMPL THERMOCOUPLE_TYPE_E ZCMPI(MEM2, "E") 1.021 HEAD {T-type THERMOCOUPLE} 1.022 5500 -100.0degC 0.15% 0.7U _T TC 2W 2.001 5500 -100.0degC _T TC S 2W 2.002 EVAL Is the CH[MEM] recording at the {left edge of chart +/-1div}? 3.001 5500 150.0degC 0.15% 0.7U _T TC 2W 4.001 5500 150.0degC _T TC S 2W 4.002 EVAL Is the CH[MEM] recording at the {center of chart +/-1div}? 5.001 5500 400.0degC 0.15% 0.7U _T TC 2W 6.001 5500 400.0degC _T TC S 2W 6.002 EVAL Is the CH[MEM] recording at the {right edge of chart +/-1div}? 7.001 JMP 47.001 7.002 LABEL THERMOCOUPLE_TYPE_R 7.003 HEAD {R-type THERMOCOUPLE} 7.004 5500 0.0degC 3.7U _R TC 2W 8.001 5500 0.0degC _R TC S 2W 8.002 EVAL Is the CH[MEM] recording at the {left edge of chart +/-1div}? 9.001 5500 500.0degC 0.15% 1U _R TC 2W 10.001 5500 500.0degC _R TC S 2W 10.002 EVAL Is the CH[MEM] recording at the {center of chart +/-1div}? 11.001 5500 1000.0degC 0.15% 1U _R TC 2W 12.001 5500 1000.0degC _R TC S 2W 12.002 EVAL Is the CH[MEM] recording at the {right edge of chart +/-1div}? 13.001 JMP 47.001 13.002 LABEL THERMOCOUPLE_TYPE_S 13.003 HEAD {S-type THERMOCOUPLE} 13.004 5500 0.0degC 3.7U _S TC 2W 14.001 5500 0.0degC _S TC S 2W 14.002 EVAL Is the CH[MEM] recording at the {left edge of chart +/-1div}? 15.001 5500 500.0degC 0.15% 1U _S TC 2W 16.001 5500 500.0degC _S TC S 2W 16.002 EVAL Is the CH[MEM] recording at the {center of chart +/-1div}? 17.001 5500 1000.0degC 0.15% 1U _S TC 2W 18.001 5500 1000.0degC _S TC S 2W 18.002 EVAL Is the CH[MEM] recording at the {right edge of chart +/-1div}? 19.001 JMP 47.001 # B Type TC only tested at 600degC and 1000degC 19.002 LABEL THERMOCOUPLE_TYPE_B 19.003 HEAD {B-type THERMOCOUPLE} 19.004 5500 600.0degC 2U _B TC 2W 20.001 5500 600.0degC _B TC S 2W 20.002 EVAL Is the CH[MEM] recording at {60% on the chart +/-1div}? 21.001 5500 1000.0degC 0.15% 0.5U _B TC 2W 22.001 5500 1000.0degC _B TC S 2W 22.002 EVAL Is the CH[MEM] recording at the {right edge of chart +/-1div}? 23.001 JMP 47.001 23.002 LABEL THERMOCOUPLE_TYPE_K 23.003 HEAD {K-type THERMOCOUPLE} 23.004 5500 0.0degC 0.15% 0.7U _K TC 2W 24.001 5500 0.0degC _K TC S 2W 24.002 EVAL Is the CH[MEM] recording at the {left edge of chart +/-1div}? 25.001 5500 500.0degC 0.15% 0.7U _K TC 2W 26.001 5500 500.0degC _K TC S 2W 26.002 EVAL Is the CH[MEM] recording at the {center of chart +/-1div}? 27.001 5500 1000.0degC 0.15% 0.7U _K TC 2W 28.001 5500 1000.0degC _K TC S 2W 28.002 EVAL Is the CH[MEM] recording at the {right edge of chart +/-1div}? 29.001 JMP 47.001 29.002 LABEL THERMOCOUPLE_TYPE_J 29.003 HEAD {J-type THERMOCOUPLE} 29.004 5500 0.0degC 0.15% 0.7U _J TC 2W 30.001 5500 0.0degC _J TC S 2W 30.002 EVAL Is the CH[MEM] recording at the {left edge of chart +/-1div}? 31.001 5500 500.0degC 0.15% 0.5U _J TC 2W 32.001 5500 500.0degC _J TC S 2W 32.002 EVAL Is the CH[MEM] recording at the {center of chart +/-1div}? 33.001 5500 1000.0degC 0.15% 0.5U _J TC 2W 34.001 5500 1000.0degC _J TC S 2W 34.002 EVAL Is the CH[MEM] recording at the {right edge of chart +/-1div}? 35.001 JMP 47.001 35.002 LABEL THERMOCOUPLE_TYPE_N 35.003 HEAD {N-type THERMOCOUPLE} 35.004 5500 0.0degC 0.15% 0.7U _N TC 2W 36.001 5500 0.0degC _N TC S 2W 36.002 EVAL Is the CH[MEM] recording at the {left edge of chart +/-1div}? 37.001 5500 500.0degC 0.15% 0.7U _N TC 2W 38.001 5500 500.0degC _N TC S 2W 38.002 EVAL Is the CH[MEM] recording at the {center of chart +/-1div}? 39.001 5500 1000.0degC 0.15% 0.7U _N TC 2W 40.001 5500 1000.0degC _N TC S 2W 40.002 EVAL Is the CH[MEM] recording at the {right edge of chart +/-1div}? 41.001 JMP 47.001 41.002 LABEL THERMOCOUPLE_TYPE_E 41.003 HEAD {E-type THERMOCOUPLE} 41.004 5500 -100.0degC 0.15% 0.5U _E TC 2W 42.001 5500 -100.0degC _E TC S 2W 42.002 EVAL Is the CH[MEM] recording at the {left edge of chart +/-1div}? 43.001 5500 150.0degC 0.15% 0.5U _E TC 2W 44.001 5500 150.0degC _E TC S 2W 44.002 EVAL Is the CH[MEM] recording at the {center of chart +/-1div}? 45.001 5500 400.0degC 0.15% 0.5U _E TC 2W 46.001 5500 400.0degC _E TC S 2W 46.002 EVAL Is the CH[MEM] recording at the {right edge of chart +/-1div}? 47.001 END